GenAlex Medical, a leading manufacturer of medical laboratory equipment, is designing a new automated system that can detect borderline high levels of dissolved cholesterol (to), using a blood sample that is as small as. Calculate the minimum mass in milligrams of cholesterol that the new system must be able to detect. Be sure your answer has the correct number of significant digits.

+4
Answers (1)
  1. 23 April, 05:48
    0
    The minimum mass of dissolved cholesterol that the new system must be able to detect is 0.43 milligrams.

    Note: The question is incomplete. The complete question is:

    GenAlex Medical, a leading manufacturer of medical laboratory equipment, is designing a new automated system that can detect borderline high levels of dissolved cholesterol (170, to 200. mg/dL), using a blood sample that is as small as 250 μL Calculate the minimum mass in milligrams of cholesterol that the new system must be able to detect Be sure your answer has the correct number of significant digits.

    Explanation:

    Range of the equipment to detect cholesterol in blood = 170.0 to 200.0 mg/dL

    Minimum amount of cholesterol that can be detected by instrument = 170 mg/dL

    Volume of blood sample = 250 μL = 0.0025 dL (1 dL = 100000 μL)

    Amount of dissolved cholesterol in 250 μL sample of blood =

    170 mg/dL * 0.0025 dL = 0.425 mg

    The minimum mass of cholesterol that the new system must be able to detect is 0.43 milligrams.
Know the Answer?
Not Sure About the Answer?
Get an answer to your question ✅ “GenAlex Medical, a leading manufacturer of medical laboratory equipment, is designing a new automated system that can detect borderline ...” in 📙 Biology if there is no answer or all answers are wrong, use a search bar and try to find the answer among similar questions.