16 December, 00:54

A calibration curve was created to determine the quantity of protein in a solution. The calibration curve has the form of a straight line with the equation A = 0.0182 x + 0.007 where A is the corrected absorbance of the solution and x is quantity of protein in the solution in units of micrograms (μg). Determine the quantity of protein in a solution that has an absorbance of 0.234. A blank solution has an absorbance of 0.055.

+4
Answers (1)
  1. 16 December, 03:53
    0
    12.5 µg

    Explanation:

    We are given the absorbance from the calibration curve:

    A = 0.0182 x + 0.007

    A = 0.234 (It is already corrected, do not need the blank value)

    so we can solve for x, the quantity of protein

    x = (0.234 - 0.007) µg / 0.0182 = 12.5 µg

    Note we are given the value for a blank solution but we do not need it for the absorbance of our sample since it is already corrected as stated in the question. If it were not a corrected absorbance, we will need to correct the absorbance value. The absorbance of the blank is caused by some absorption from the solvent, cuvette, internal reflection, etc.
Know the Answer?
Not Sure About the Answer?
Get an answer to your question ✅ “A calibration curve was created to determine the quantity of protein in a solution. The calibration curve has the form of a straight line ...” in 📙 Chemistry if there is no answer or all answers are wrong, use a search bar and try to find the answer among similar questions.